Non-contact Quantitative Diffuse Reflectance Spectroscopy
V. Sorgato
,
M. Berger
,
C. Emain
,
A. Koenig
,
B. Roig
,
C. Vever-Bizet
,
J.M. Dinten
,
G. Bourg-Heckly
,
A. Planat-Chretien
DIFFUSE OPTICAL IMAGING V, 9538, 95380U (2015)
DOI: 10.1117/12.2183764
ISSN: 0277-786X
Abstract
We validate a non-contact Diffuse Reflectance Spectroscopy (DRS) system as a first stage to approach quantitative multi-spectral imaging technique. The non-contact DRS system with separated illumination and detection paths was developed with different progressive set-ups which were all compared to a well-founded contact DRS system. While quantitation of the absorption coefficient is well achieved with the existing method, the calculation of the scattering coefficient is deteriorated by the non-contact architecture measurements. We have therefore developed an adaptive reference-based algorithm to compensate for this effect.